Acqiris averagers can be operated in Averager Mode or in standard Digitizer Mode.

AcqirisLive operates in either mode, but since these are mutually exclusive, only instruments that support the chosen
mode are accessible in any given mode. E.g. only averagers are accessible in Averager Mode, while both digitizers
and averagers are accessible in Digitizer Mode.

The operation mode can be switched within the application. By default, AcqirisLive starts in Digitizer Mode.

Buffer size

Upon startup of the application, the available physical memory on the computer is checked. The total data buffer size
is then limited to the available memory minus 10% of the total memory. It is also possible to limit the total data
buffer size to a fixed value with the -bn option (where n is the limit in MB, e.g. -b200 will limit the buffer to 200
MB). The data buffer is then shared among the active channels.

When the total acquisition memory (sum over all active channels) exceeds the data buffer size limit, readout of the
data will be partial. The special operation modes such as Segments Average, Average, Digital Filter and FFT are not
supported with partial data readout.

Partial readout is controlled with the Very Long Memory panel (menu Options ->Very Long Memory...). The first
segment and the first sample to read out can be specified. Note that if the buffer is large enough for at least 1 full
segment, only full segments will be read (i.e. first sample control is ignored).

When saving data to file, you will be prompted for a maximum number of segments per file.

Banks for SAR

The U1071A-FAMILY and some of the 10-bit FAMILY modules are capable of acquiring data and being read out
simultaneously. The -Bn option allows the internal memory of the digitizer to be divided into n banks. To observe
the improved acquisition trigger rate achievable in this mode it is suggested to set up the desired conditions and then
turn the event display off. The trigger out signal can be used to monitor the trigger rate; an interval width trigger can
be used to check if triggers are missing.

Calibration

Before operation, the instruments must undergo an internal calibration. As this calibration can only proceed
sequentially for each channel, the initialization time of AcqirisLive can become significant for systems with many
channels. Therefore, it can sometimes be desirable to be able to start the application quickly by skipping the
calibration (which must then be done "manually" before collecting correct data).

DMA Data Transfers

By default, data transfers on the PCI bus between the digitizer memory and the PC are done in DMA (Direct
Memory Access) mode, which allows the fastest transfer speeds (up to about 130 MB/s peak, 100 MB/s sustained
acquisition data rate). These very high data rates are not well supported by all hardware configurations (PC
motherboard), and system hangs due to endless retries on the PCI bus have been observed on some machines. Such
problems disappear when switching DMA mode off (data transfer rate is then about 2.8 MB/s).

Please note that switching DMA data transfer mode off precludes the use of acquisition interrupts. The –i switch will
therefore be ignored when –n is used.

Process Priority

When the program checks for an event from the operating system (e.g. to react to user input), it can put itself to sleep
for a specified period of time. This gives other applications more processor time, but limits the performance of
AcqirisLive in terms of maximum acquisition rate. The default is not to go to sleep, but this can be overridden to
allow better performance for other applications. Note that this sleep policy is only in effect while the acquisition is
running. When the acquisition stops, the sleep policy is always reset to the longest sleep period.

Acquisition Interrupts

When performing acquisitions, there are 2 possible ways to wait for the acquisition to complete: polling or using a
hardware interrupt.

-p Polling: AcqirisLive runs as a single thread process. After an acquisition is started, the instruments are

regularly polled to check when the acquisition is complete. The user interface is still alive, but the latency
for reading the data after the acquisition may vary.

-i Hardware interrupt: In this mode, AcqirisLive is run as a multi-thread process. After starting an

acquisition, a new thread is created to wait for the hardware interrupt indicating that the acquisition