Technical data, Specifications, Performance – Vaisala PTB330 User Manual

Page 127: Barometric pressure range 500 0 hpa, Barometric pressure range 50 0 hpa, Product, Specifications performance, Chapter 7

Advertising
background image

Chapter 7 ____________________________________________________________ Technical Data

VAISALA ______________________________________________________________________ 125

CHAPTER 7

TECHNICAL DATA

This chapter provides the technical data of the product.

Specifications

Performance

Barometric pressure range 500 ... 1100 hPa

Table 20

Barometric Pressure range 500 ... 1100 hPa at 20°C

Class A

Class B

Linearity*

±0.05 hPa

±0.10 hPa

Hysteresis*

±0.03 hPa

±0.03 hPa

Repeatability*

±0.03 hPa

±0.03 hPa

Calibration uncertainty**

±0.07 hPa

±0.15 hPa

Accuracy at +20 °C (+68 °F)*** ±0.10 hPa

±0.20 hPa

Barometric pressure range 50 ... 1100 hPa

Table 21

Barometric pressure range 50 ... 1100 hPa at 20°C

Class B

Linearity* ±0.20

hPa

Hysteresis* ±0.08

hPa

Repeatability* ±0.08

hPa

Calibration uncertainty**

±0.15 hPa

Accuracy at ±20 °C (+68 °F)*** ±0.20 hPa

Advertising