70 __________________________________________________________________ M210855EN-D

Setting and testing the analog outputs
AMODE

Displays the analog output mode (if an
AOUT-1 module is connected).

ASEL

Sets the analog output quantity and scaling
(low/high).

ACAL

Adjusts the analog output.

AERR

Sets the analog output error value.

ATEST

Starts/ends an analog output test.

Setting and testing the relays
RSEL

Sets the relay scaling (if a RELAY-1
module is connected).

RTEST

Starts/ends a relay output test.

General Settings

Changing Quantities and Units

NOTE

Only the quantities selected when ordering the device can be selected
as a display output quantity.

FORM

Use the serial line command FORM to change the format or define
quantities for the output command SEND and for RUN mode.

FORM [x]

where:

x = Formatter

string

Formatter string consists of quantities and modifiers. When selecting
the quantity, use the abbreviations of the quantities presented in Table
2 on page 15. The m
odifiers are presented in Table 12 on page 71.