Using serial line, Seri, Smode – Vaisala MMT330 User Manual

Page 111

Advertising
background image

Chapter 4 _________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 109

The new user port settings set using the display/keypad are effective
immediately.

Using Serial Line

NOTE

You can use the serial commands to change/view the user port settings
even if you are currently connected to the service port.

SERI

Use the SERI command to set the communication settings for the user
port. Changed settings will be activated at next reset or power up.

SERI [b p d s]<cr>

where

b

=

Bit rate (110, 150, 300, 600, 1200, 2400, 4800, 9600,19200,
38400, 57600, 115200)

p

=

Parity (n = none, e = even, o = odd)

d

=

Data bits (7 or 8)

s

=

Stop bits (1 or 2)

The settings can be changed one parameter at a time or all parameters at
once.

Example (changing all parameters):

>seri 600 N 8 1
600 N 8 1
>

Example (changing parity only):

>seri O

4800 O 7 1

SMODE

Use the SMODE command to set the user port start-up operating mode.

SMODE [xxxx]<cr>

where

xxx =

STOP, SEND, RUN, POLL, or MODBUS.

Advertising