Table 13, Alsel parameters – Vaisala MMT330 User Manual

Page 79

Advertising
background image

Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 77

Table 13

ALSEL Parameters

Name

Description

above

If specified, the alarm goes off when the quantity value rises above
the exact value of this setpoint. However, if above < below, the
alarm goes off at (above + hyst) and (below – hyst) instead.
Cannot be set for fault and online alarms.

below

If specified, the alarm goes off when the quantity value decreases
below the exact value of this setpoint. However, if above < below,
the alarm goes off at (above + hyst) and (below – hyst) instead.
Cannot be set for fault and online alarms.

hyst

Distance from setpoint amount of extra retractment needed in order
to regenerate alarm on the next crossing of the setpoint.
Cannot be set for fault and online alarms.

enabl

Enable (ON) or disable (OFF) the display alarm.

Example (set aw and T as display alarm quantities and configure the
alarms as prompted):

>alsel aw t
Al1 aw above: 0.50 ? 0.6
Al1 aw below: - ?
Al1 aw hyst : 0.10 ?
Al1 aw enabl: ON ?
Al2 T above: - 'C ?
Al2 T below: - 'C ? 0
Al2 T hyst : 0.00 'C ? 1
Al2 T enabl: OFF ? ON
>

Advertising