Chapter 4 _________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 127

RSEL

Use the serial line to select the quantity, setpoints and hysteresis or
enable/disable the relay outputs. Enter the RSEL command.

RSEL [q1 q2 q3 q4]<cr>

where

q1

=

Quantity for the relay 1 or Fault/Online

q2

=

Quantity for the relay 2 or Fault/Online

q3

=

Quantity for the relay 3 or Fault/Online

q4

=

Quantity for the relay 4 or Fault/Online

Factory setting: All relays disabled.

The MMT330 measures the following quantities:

- Water activity (aw)

- Temperature (T) (metric unit: ºC, non-metric unit:ºF)

- ppm for transformer oil only (H

2

O) (optional quantity)

Example of window limit switch: Selecting relay 1 to follow aw
measurement and relay 2 to follow temperature measurement. Two relay
setpoints are set for both relays.

>rsel aw t
Rel1 aw above: 0.00 ? 0.3
Rel1 aw below: 0.00 ? 0.4
Rel1 aw hyst : 0.00 ? 0.02
Rel1 aw enabl: OFF ? on
Rel2 T above: 0.00 'C ? 30
Rel2 T below: 0.00 'C ? 40
Rel2 T hyst : 0.00 'C ? 3
Rel2 T enabl: OFF ? on