Aalborg Totalizer User Manual

Page 9

Advertising
background image

6

TABLE 1

1

If Totalizer has label with
corresponding flow rate on the back
and you do not plan to change the
full scale range.

The Totalizer is ready for use and
does not require any adjustment.

2

If Totalizer has label with
corresponding flow rate on the
back and you do plan to change
the full scale range.

Refer to f.4 RE-SCALING OF
TOTALIZER

3

If Totalizer does not have label
with corresponding flow rate on the
back (default calibration for 30
unit/min) and you do plan to change
the full scale range.

Refer to f.4 RE-SCALING OF
TOTALIZER

4

ƽ Use this option only if the
potentiometers R12 (GAIN) or R19
(OFFSET) were accidentally
readjusted.

1. Refer to f.3 and perform

GAIN ADJUSTMENT

2. Refer to f.4 and perform

RE-SCALING OF TOTALIZER for
desired flow range.

Choose from TABLE 1 the option which better matches your case: