Aalborg Vortex User Manual

Page 100

Advertising
background image

95

COMMAND

NAME

DESCRIPTION

NO.

COMMAND SYNT

A

X

COMMAND

ARGUMENT 1

ARGUMENT 2

ARGUMENT 3

ARGUMENT 4

RESPONSE

4-20mA

Analog

Output

Settings

Sets/Reads Analog Outputs assignment,

4mA and 20mA V

alues. NOTE: 4mA and

20mA values have to be entered in

corresponding to Argument 2 type EU

(for V

o

lum. in %FS). 0 - Disabled.

1-V

o

lum. Flow: Current active

volumetric flow EU. 2-Mass Flow:

Current active volumetric flow EU.

3-T

emperature: deg. C 4-Pressure: Pa.

22

AO

<V

alue>

Channel: 1 or

2

<V

alue> Mode: [0-4]

0 - Disabled

1-V

o

l. Flow

2-Mass Flow

3-T

emp.**

4-Pressure**

<V

alue>

4mA value

(in EU related

to Mode

value)

<V

alue>

20mA value

(in EU related

to Mode

value)

Ch1

AO1:0,0,100

Ch2

AO2:2,0.0,26.0

No Argument (Ret.

Current settings. For

V

o

lum. Flow in %FS)

Ch1. AO1:1,0.0,1000.0

Ch2. AO2:2,0.0,26.0

Data Log

Reads Data Log entries, Sets

Mode and Configuration of

the data logger

. Argument 1 = C

DL Configuration (Argument2):

0 – Loop Enabled

1 – Loop Disabled

2 – Loop Disabled,

number of samples to record DL

Configuration (Argument3): Number of

samples: [1-15871] (has to be

submitted for all options, but only

applicable for Argument 2 value 2).

Argument 1 = V

. Process V

ariable:

0-V

o

lumetric Flow 1-Mass Flow

2-T

emperature**. 3-Pressure**

Argument 1 = T

0 – Unconditional

1 – PV Alarm High Event

2 – PV Alarm Low Event

23

DL

M

Mode:

E – DL Enable

D – DL Disable

<V

alue>

E or D

DLM:<value>

Example: DLM:E

No Argument (Ret.

Current settings)

DLM:<value>

Example: DLM:E

C

Configuration

<V

alue> [0-2]

<V

alue>

: [1-15871]

DLC:<value>,<value>

Example: DLC:0,1024

No Argument (Ret.

Current settings)

DLC:<value>,<value>

Example: DLC:0,1024

V

Process

V

ariable

<V

alue> [0-3]

DL

V

:<value> Example: DL

V

:0

No Argument (Ret.

Current settings)

DL

V

:<value>

Example: DL

V

:0

I

Sample

Inter

val in sec.

<V

alue> [1-3600]

DLI:<value> Example: DLI:1

(Ret. Current

settings)

DLI:<value> Example: DLI:1

T

DL Start/Stop

conditions

Thresholds

<V

alue>

Start Cond[0-2]

<V

alue>

Stop: [0-2]

DL

T:<value>,<value>

Example: DL

T:0,0

(Ret. Current

settings)

DL

T:<value>,<value>

Example: DL

T:0,0