SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES

7-4

PROGRAM

01:

P17

Panel Temperature

01:

1

IN Card

02:

1

Loc [:PANL TEMP]

02:

P13

Thermocouple Temp (SE)

01:

5

Reps

02:

2

5000 uV slow Range

03:

1

IN Card

04:

2

IN Chan

05:

22

Type E (Skip every other chan)

06:

1

Ref Temp Loc PANL TEMP

07:

2

Loc [:S.E. T#1 ]

08:

1

Mult

09:

0

Offset

03:

P14

Thermocouple Temp (DIFF)

01:

5

Reps

02:

1

1500 uV slow Range

03:

1

IN Card

04:

1

IN Chan

05:

12

Type E (Temp difference)

06:

2--

Ref Temp Loc S.E. T#1

07:

7

Loc [:DIFF T #1]

08:

1

Mult

09:

0

Offset

When the temperature of the R junction is
outside of the CR7 reference junction
compensation range (Table 13.4-3), the TCs
must be connected in the normal fashion, one
TC per input channel; both temperatures
measured and one subtracted from the other to
find the difference. This must be done because
any error in the reference junction
compensation becomes an error in the
temperature difference.

7.6 TEMPERATURE WITH CALIBRATED

THERMOCOUPLES

Thermocouple calibration (Section 13.4) results
in a slope correction. The correction must be
applied only to the thermocouple output. When
Instructions 13 and 14 are used to measure
temperature, the temperature is the sum of the
reference temperature and the temperature
difference calculated from the thermocouple
output. The correction must be applied to the
temperature difference before the reference
temperature is added.

Example A demonstrates the use of a scaling
array (Instruction 53) to correct the calibration of
four individually calibrated thermocouples.

Another means of applying a correction factor to
a number of thermocouples is to group together
those with a similar correction factor. In
example B, the slope correction factor for a
group of 5 thermocouples is entered as the
multiplier (Parameter 8) in the instruction to
read those thermocouples. The example only
shows one group of thermocouples. If there
were several groups with similar correction
factors, Instruction 14 would be used to read
and correct each group.

After the slope correction is made, a loop is
used to add the reference temperature to the
corrected temperature differences.

CONNECTIONS

The thermocouples are connected in the normal
manner: chromel to Hi and constantan to Low.
In both examples, the first thermocouple is
connected to Channel 1. Care must be taken
that the correction factors called for in the
programming match the channels that the
calibrated thermocouples are connected to.

PROGRAM A

01:

P17

Panel Temperature

01:

1

IN Card

02:

1

Loc [:REF TEMP ]

02:

P14

Thermocouple Temp (DIFF)

01:

4

Reps

02:

3

15 mV slow Range

03:

1

IN Card

04:

1

IN Chan

05:

12

Type E (Temp difference)

06:

1

Ref Temp Loc REF TEMP

07:

2

Loc [:TC temp#1]

08:

1

Mult

09:

0

Offset

03:

P53

Scaling Array (A*loc +B)

01:

2

Start Loc [:TC temp#1]

02:

.99255 A1

03:

0

B1

04:

.99703 A2

05:

0

B2

06:

1.0045

A3

07:

0

B3

08:

1.0075

A4

09:

0

B4