TD ADDENDUM—OVERVIEW

AD-OV-4

Key

(ID:Data)

Explanation

*0

LOG 1

Exit Table 1, enter *0 Mode, compile table and begin
logging.

*6

06:0000

Enter *6 Mode (to view Input Storage).

A

01:21.234

Advance to first storage location. Panel temperature is
21.234

°C (the display will show the actual

temperature).

Wait a few seconds:

01:21.423

The CR10X has read the sensor and stored the result
again. The internal temp is now 21.423

°C. The value

is updated every 5 seconds when the table is executed.
At this point the CR10X is measuring the temperature
every 5 seconds and sending the value to Input
Storage. No data are being saved. The next step is to
have the CR10X send each reading to Final Storage.

*1

01:00

Exit *6 Mode. Enter program table 1.

2A

02:P00

Advance to 2nd instruction location (this is where we
left off).

84

02:P84

This is the Data Table Instruction.

A

01:0.0000

Enter 84 and advance to the first parameter (which is
the time into the interval).

0

01:0

This parameter determines when in the output internal
data is stored. 0 stores data on the even interval.

A

02:0.000

Enter 0 and advance to the second parameter.

0

02:0

This parameter specifies the output interval. 0 stores
data each execution.

A

03:0.0000

Enter 0 and advance to third parameter.

1000

03:1000.00

This parameter specifies how many records to store in
the table before overwriting the oldest. For this
example, keep 1000 records.

A

03:P00

Enter 1000 and advance to the third program
instruction.

70

03:P70

The SAMPLE instruction. It directs the CR10X to take
a reading from an Input Storage location and send it to
Final Storage (an Output Processing Instruction).

A

01:0000

Enter 70 and advance to the first parameter
(repetitions).

1

01:1

There is only one input location to sample; repetitions = 1.

A

02:0000

Enter 1 and advance to second parameter (Input
Storage location to sample).

1

02:1

Input Storage Location 1, where the temperature is
stored.

A

04:P00

Enter 1 and advance to fourth program instruction.

*

00:00

Exit Table 1.

0

LOG 1

Enter *0 Mode, compile program, log data.