Theory of Operation

Functionality of the System

Figure 2-2

System functionality block diagram for U2802A

13 Thermocouple voltage signals are detected at the U2802A thermocouple

inputs.

14 Signal is amplified with a gain of 97.673 by the U2802A.

15 The U2355A or U2356A DAQ converts the analog voltage signals to

digital voltage readings.

16 The AMM software (or IVI- COM driver) reads the Gain and Offset

calibration constants from the U2802A EEPROM via the DAQ DIO lines.
The digital voltage readings will be calibrated based on these constants.

17 The AMM software (or IVI- COM driver) converts the calibrated voltage

readings to temperature readings using the ITS- 90 conversion
polynomials.

U2802A_UG.book Page 21 Friday, November 25, 2011 10:57 AM