Table 4- 3 – Agilent Technologies U2802A User Manual

Page 55

Advertising
background image

Product Specifications

4

Agilent U2802A 31-Channel Thermocouple Input User’s Guide

45

Table 4-3 U2802A measurement accuracy with U2356A, at 0 to 18 °C and 28 to 45 °C,

with different number of averaging points.

Thermocouple Measurements Accuracy

(U2356A @ 0 °C to 18 °C and 28 °C to 45 °C)

T/C

Type

ITS-90

Temperature

Range (°C)

Optimum

Measurement

Range(°C)

Without

averaging

(± °C)

50 points

averaging

(± °C)

500 points

averaging

(± °C)

Low

High

Low

High

B

0

1820

1100

1820

6.1

3.1

2.4

400

1100

14.4

6.3

2.7

E

–270

1000

–150

1000

3.0

2.6

2.6

–200

–150

4.2

3.7

3.6

J

–210

1200

–150

1200

2.9

2.5

2.5

–210

–150

4.9

4.1

4.0

K

–270

1372

–100

1200

3.3

2.9

2.9

–200

–100

5.3

4.2

4.0

N

–270

1300

–100

1300

3.4

2.7

2.6

–200

–100

6.8

4.6

4.1

R

–50

1768

300

1760

6.2

3.7

3.2

–50

300

15.7

7.2

3.8

S

–50

1768

400

1760

6.4

4.0

3.4

–50

400

14.2

6.6

3.4

T

–270

400

–100

400

3.0

2.4

2.2

–200

–100

5.3

4.2

3.9

U2802A_UG.book Page 45 Friday, November 25, 2011 10:57 AM

Advertising