Product characteristics – Agilent Technologies U2802A User Manual

Page 51

Advertising
background image

Product Specifications

4

Agilent U2802A 31-Channel Thermocouple Input User’s Guide

41

Product Characteristics

GENERAL CHARACTERISTICS

Number of channels

31 differential and 1 CJC

Input voltage range for voltage mode

±10 V (signal + common mode)

Input voltage (thermocouple mode)

±100 mV

Sampling rate for thermocouple mode

10 kSa/s total for all channels

Sampling rate for overall module

500 kSa/s

Thermocouple types

J, K, R, S, T, N, E and B

INPUT SPECIFICATIONS

Accuracy (thermocouple mode)

Overall gain error

Overall offset error

Nonlinearity

0.06% (23 °C ± 5 °C)

15 µV (without zeroing) (23 °C ± 5 °C)

6 µV (with zeroing)

< 0.005% of Full Scale Range

System noise (rms)

Gain (x1)

Gain (x100)

100 µVrms

5 µVrms

Common mode rejection ratio (60 Hz)

Voltage mode

Thermocouple mode

> 60 dB

> 80 dB

Cold junction accuracy

±1.0 °C typical (23 °C ± 5°C)

±1.5 °C typical (0 °C to 18 °C, 28 °C to 55 °C)

INPUT CHARACTERISTICS

Bandwidth (voltage mode)

> 500 kHz

Bandwidth (thermocouple mode)

4.0 Hz

U2802A_UG.book Page 41 Friday, November 25, 2011 10:57 AM