A&D Automatic Blood Pressure Monitor TM-2655P User Manual

Page 16

Advertising
background image

8. When the measurement is complete, the air is automatically

released from the cuff and the cuff will become loose.

9. The measurement results appear in the display.

10. (TM-2655P) The measurement results are output to the

printer.

Fig.9 Measurement results

Note

To stop the measurement at any time, press the START/STOP switch. The quick exhaust will

reduce the pressure in the cuff and the cuff will become loose.

If the air is not exhausted even when the START/STOP switch is pressed, press the

EMERGENCY STOP switch.

If your arm can not be removed from the cuff due to the power failure, press the safety lever

located on the front lower part of the device. It disconnects the clutch to loosen the cuff.

Please wait for about 10 minutes before repeating the measurement.

10

Advertising
This manual is related to the following products: