Communication specifications, 1 channel 1 : miniature din 8-pin – A&D Automatic Blood Pressure Monitor TM-2655P User Manual

Page 23

Advertising
background image

8. COMMUNICATION SPECIFICATIONS

The TM-2655/TM-2655P is equipped with two RS-232C channels. Various settings for each channel are
available in the function setting mode. Refer to “7-2 Description of the Functions” for details.

8-1 Channel 1 : Miniature DIN 8-pin

CAUTION

The personal computer and medical equipment connected to the device are not allowed to

be in the patient area.

The personal computer used must conform to IEC60950

Specifications

Table 6 Channel 1 specifications

Standard

Conforms to EIA RS-232C

Transmission Asynchronous,

half-duplex

Baud rate

1200, 2400, 4800, 9600 bps (Can be changed in the function setting mode “F07”.)

Data bits

8 bits

Parity bit

None

Stop bits

1 bit

Code ASCII

Equipment available for connection

Channel 1:

Scales, and automatic weight and height scales manufactured by A&D

Channel 2:

Personal computer

Pin assignment

Pin No.

Signal name

Description

1 TXD

Send

data

2 RXD

Receive

data

3

RTS

Ready to send

4 -

Not

connected

5

CTS

Clear to send

6 GND

Signal

ground

7 -

Not

connected

8 -

Not

connected

Note

Do not use pins 4, 7, and 8.
They are used by the device.

Fig.14 Pin assignment

Cable connection between the device and a personal computer

TM-2655/TM-2655P Personal computer

Fig.15 Cable connection diagram