9-2 Replacing the Fuses

Replace the fuses as follows:

1. Remove the caps of both fuse holders located on the rear

panel.

2. Replace both fuses with new ones.

3. Re-install the fuse holder caps.

Fig.19 Fuse holders

Note

Use only fuses of the proper type, voltage and current rating as specified on the rear panel, and that
conform to IEC60127.

9-3 Checking the Counter

The TM-2655/TM-2655P is equipped with a counter function, which indicates how many times the device

has measured blood pressure. The counter data remains in memory even after the power is turned off.

Use the counter function to check the frequency of the device use or as the guideline for a periodic

cleaning

Displaying the counter

Press the COUNTER switch. The counter data appears in the

SYSTOLIC and DIASTOLIC displays for about 4 seconds.

Note

Do not keep the COUNTER switch pressed. It will reset the counter
data to zero.

Resetting the counter

Fig.20 COUNTER switch

Hold down the COUNTER switch for 4 seconds or more. The

counter data will be reset to zero.

9-4 Error Codes

Table 9 Error codes

Error code

Description

pe

The printer paper has run out. Install a new roll of printer paper.

HU

The printer head is up. Lower the lever and lock the printer paper in position.

err

A measurement error has occurred. Refer to “10. Troubleshooting”.