Installation – A&D Automatic Blood Pressure Monitor TM-2655P User Manual

Page 14

Advertising
background image

3. INSTALLATION

Follow the procedure below to install the TM-2655/TM-2655P.

1. To ensure that the TM-2655/TM-2655P works properly, install the device in an environment where:

The temperature range is from 10

°C to 40°C (50°F to 104°F).

The humidity is less than 85% (non condensing).

It is away from water, dust, chemicals, and corrosive or explosive gases.

It is not exposed to direct sunlight.

2. Place the TM-2655/TM-2655P on a table solid enough to support its weight.

3. Adjust the height of the chair and table so that the arm insertion section is at the user’s heart level.

Fig.4 Proper installation

4. Use the power cable provided with the device to connect between the AC INPUT connector and an

electrical outlet.

Using the anti-theft opening, secure the device to the table with a solid chain.

Fig.5 Power cable connection

Note

Be sure to use the correct voltage for the power source.

8

Advertising