background image

4. OPERATION

1. Turn on the POWER switch located on the rear panel.

When the POWER switch is turned on, all of the display

symbols will appear for several seconds. Then;

(TM-2655) “0” (zero) appears in the SYSTOLIC pressure

display, indicating that the device is ready for
measurement.

(TM-2655P) The human sensor functions for three

minutes. When it does not detect a person, “ .
“ (dot) appears in the SYSTOLIC pressure
display, indicating that the device is ready for
measurement.

Fig.6 POWER switch and
SYSTOLIC display (TM-2566)

2. Take off your jacket.

If you wear a thick jacket, take it off for a better measurement.

Note

Wearing a thick jacket may cause a faint pulse, and result in a
measurement error.

3. Adjust the height of the chair and table.

Adjust the height of the chair and table so that the arm

insertion section is at your heart level.

4. Insert your arm into the arm insertion section.

Note

Fig.7 Proper posture

Be sure to insert your arm to the shoulder.

5. Press the START/STOP switch to start measurement.

6. The cuff will be pressurized automatically up to the

predetermined pressure. Refer to “7-2 Description of the
Functions” for details.

7. When pressurization is complete, the automatic exhaust

mechanism will gradually reduce the pressure in the cuff. Just
relax and remain still.

Fig.8 Measuring procedure

9