Agilent Technologies Agilent 4396B User Manual

Page 90

Advertising
background image

3

dB

Bandwidth

The

analyzer

calculates

the

bandwidth

of

the

DUT

between

two

equal

power

levels

.

In

this

example

,

it

calculates

the

03

dB

bandwidth

relative

to

the

lter

center

frequency

.

1.

Press

4

Ma

rk

er

5 .

Then

use

the

rotary

knob

to

move

the

marker

to

the

center

of

the

lter

passband.

2.

Press

4

Search

5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTH

[OFF]

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTH

on

OFF

to

NNNNNNNNNNNNNNNNNNNN

ON

off

.

The

analyzer

calculates

the

03

dB

bandwidth,

center

frequency

,

Q

(Quality

F

actor),

insertion

loss

,

and

di erences

between

the

center

frequency

and

the

cuto

frequencies

of

the

DUT

.

It

then

lists

the

results

at

the

upper

right

hand

of

the

display

.

Sub-marker

1

on

the

trace

shows

the

passband

center

frequency

and

sub-markers

2

and

3

show

the

location

of

the

03

dB

cuto

points

.

Figure

5-3.

Using

the

Marker

to

Determine

3

dB

Bandwidth

T

o

have

the

analyzer

calculate

the

bandwidth

between

other

power

levels

,

select

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTH

VALUE

and

enter

the

number

(for

example

,

enter

06

4

21

5

for

06

dB).

Press

4

Ma

rk

er

5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRESET

MKRS

when

you

are

nished

with

this

measurement.

5-4

Network

Measurement

Examples

Advertising