Turning off the solvent tracking feature, Chapter 5 – Agilent Technologies 1100 Series User Manual

Page 23

Advertising
background image

Thermo Scientific

Agilent 1100 Series LC Devices Connecting the Hardware and Triggering Data Acquisition

15

5

Turning Off the Solvent Tracking Feature

The solvent tracking feature on the Agilent 1100 LC pumps is not supported by Xcalibur at
this time. You must turn off this feature to prevent error messages from terminating the data
acquisition.

Y

To turn off the solvent tracking feature on the Agilent 1100 pump

1. If Xcalibur is open, close it.

2. Open the Analysis screen on the Agilent 1100 handheld control module.

3. Press the Settings button F1.

Figure 8.

Agilent 1100 handheld control module

4. Press the number corresponding to the pump.

5. Press the Bottle Fillings button F4.

6. Set all of the solvent Total boxes to 0.

The solvent Actual boxes are then automatically set to 0.

7. Clear the Error If Empty check box by using the right arrow to move the focus to this