A&D Check Weighing Scales FS-6Ki User Manual

Page 17

Advertising
background image

15

3. Press the ENT key to store the HI limit weight in memory.

4. The display shows 5tored for a few seconds and

returns to weighing mode.

5. Press the LO key. Then, the last LO limit weight

entered will flash and the left part of analog display
turns on.

6. Enter the LO limit weight via the 10-key pad.

‰ If you hit the wrong key, press the C key and the

LO limit weight is re-displayed. Continue data entry
again.

‰ You cannot enter the “+” or “-” sign for the LO limit.

7. Press the ENT key to store the LO limit weight in memory.

8. The display shows 5tored for a few seconds and

returns to weighing mode.

‰ Example:

Target weight: 3.000 kg, HI limit weight: 0.050 kg, LO limit weight: 0.030 kg

Then,

Upper limit = 3.000 kg + 0.050 kg = 3.050 kg
Lower limit = 3.000 kg - 0.030 kg = 2.970 kg

LO: weight < 2.970 kg
OK: 2.970 kg

” weight ” 3.050 kg

HI: 3.050 kg < weight

Setting the HI limit % and LO limit %

‰ Set the Function setting “F7-2 Target weight setting mode with HI/LO limit %”.

1. Press the HI key. Then, the last HI limit % entered

will flash and the right part of analog display turns on.

2. Enter the HI limit % via the 10-key pad.

‰ If you hit the wrong key, press the C key and the

old HI limit % is re-displayed. Continue data entry
again.

‰ You cannot enter the “+” or “-” sign for the HI limit.

3. Press the ENT key to store the HI limit % in memory.

4. The display shows 5tored for a few seconds and

returns to weighing mode.

-

5tored

-

000.040

-

5tored

-0

001.50

-

5tored

-

000.030

-0

001.00

Advertising