A&D Check Weighing Scales FS-6Ki User Manual

Page 19

Advertising
background image

17

3. Press the ENT key to store the HI limit weight in memory.

4. The display shows 5tored for a few seconds and

returns to weighing mode.

5. Press the LO key and the last LO limit weight

entered will flash.

6. Enter the LO limit weight via the 10-key pad.

‰ If you hit the wrong key, press the C key and the LO

limit weight is re-displayed. Continue data entry again.

‰ You can enter minus value using the ± key. This

key toggles the sign.

7. Press the ENT key to set the LO limit weight in memory.

8. The display shows 5tored for a few seconds and

returns to weighing mode.

‰ Example:

Upper (HI) limit weight: 3.050 kg, Lower (LO) limit weight: 2.950 kg

Then,
LO: weight < 2.950 kg
OK: 2.950 kg

” weight ” 3.050 kg

HI: 3.050 kg < weight

8-3. The comparator memory

The FS-i scale can store up to 100 comparator limits by 2 digit numbers, form 00 to 99.

‰ To use this function, the scale must not be set to the simplified operation mode.

Storing a set of limits in memory

1. Set the comparator data according to the previous sections “8-1. Target weight

setting mode” (target weight, HI limit and LO limit) or “8-2. Upper & Lower limits
weight setting mode” (HI limit and LO limit).

2. Press the STORE key. Then the display will

show 5d 00 .

3. Press the memory number, for example 6.

‰ If you stop to store, press the C key, the scale

returns to the weighing mode.

4. Press the ENT key to store the data to into the

memory. The display shows 5tored for a few
seconds and returns to weighing mode.

-

5tored

-

003.560

-

002.950

-

5tored

-f

5d

-

00

-f

5d

-

06

-

5tored

Advertising