88 cal 0 f 8, 11hi00 – A&D Counting Scale HC-15Ki User Manual

Page 30

Advertising
background image

28

If you know the exact weight value, or
if you wish to use a different weight,
use the 0

Æ 9 and . 10-key pad to

display the desired calibration weight.
(For example: Using 5 kg calibration weight
that actually weighs 5.001 kg. Do not forget
to enter a decimal point in this case.)

3. Press the ENTER key.

The calibration weight stops blinking.

4. Making sure that there is nothing on, or touching

the weighing pan, press the ENTER key.
When zero calibration is completed, the display
will show “Cal f”.

If you don’t need span calibration, press the

ON/OFF key to exit from the calibration

procedure.

5. Place the calibration weight on the weighing pan

and press the ENTER key.
When span calibration is completed, the display
returns to step 1 showing the weight value for the
calibration weight.
Remove the calibration weight.

If the calibration weight is not what it should be,
an error will be displayed. Check if the weight is
correct and try again.

6. Press the ON/OFF key to turn the scale off

and re-attach the calibration switch cover.
(End of the calibration procedure.)

If the scale will be moved to another place, set the gravity
acceleration value before calibration. The value must be of the
area where the calibration is to be done.

888

Cal

10

COUNT

WEIGHT UNIT WEIGHT

88

Cal

0

0

8

8

0.0

5.001

0

00

888

Cal

10

COUNT

WEIGHT UNIT WEIGHT

88

Cal

0

0

8

8

0.0

5.000

0

00

ENTER

888

Cal

10

COUNT

WEIGHT UNIT WEIGHT

88

Cal

0

f

8

8

0.0

5.000

0

00

Nothing on the
weighing pan!

ENTER

888

Cal

02

COUNT

WEIGHT UNIT WEIGHT

88-

5.000

8

8

11Hi00

Place
Calibration

i ht

ENTER

Advertising