Myron L 6P (includes Addendum 10-01) User Manual

Page 16

Advertising
background image

XII.

TROUBLESHOOTING CHART

27

Symptom

Possible Cause

Corrective Action

No

display

, even though

Battery weak or not connected.

Check connections or replace battery

measurement key pressed

(ref. Battery Replacement, pg. 24).

Inaccurate

pH

readings

1. pH calibration needed (ref. pH Cal., pg. 17). 1. Recalibrate instrument.
2. Cross-contamination from residual pH

2. Thoroughly rinse sensor well.

buffers or samples in sensor well.

3

. Recalibrate using fresh buffers (ref. pH

3. Calibration with expired pH buffers.

Buffer Solutions, pg. 28).

No response to

pH

changes

Sensor bulb is cracked or an

Replace pH/ORP sensor (ref. pH/ORP

electromechanical short caused by an

Sensor Replacement, pg. 29).

internal crack.

Will not adjust down to

pH

7

pH/ORP sensor has lost KCl.

Clean and rejuvenate sensor (ref. Cleaning
Sensors, pg. 25) and recalibrate. If no
improvement, replace pH/ORP sensor
(ref. pH/ORP Sensor Replacement, pg. 29).

pH

readings drift or respond

1. Temporary condition due to “memory” Clean and rejuvenate sensor (ref. Cleaning

slowly to changes in

of solution in pH sensor well for long

Sensors, pg. 25) and recalibrate. If no

buffers/samples or “FAC” is

periods.

improvement, replace pH/ORP sensor

displayed repeatedly

2. Bulb dirty or dried out.

(ref. pH/ORP Sensor Replacement, pg. 29).

3. Reference junction clogged or coated.

Unstable

1. Dirty electrodes.

1. Clean cell cup and electrodes (ref.

Conductivity

/

T D S

/

2. Test samples greater than 1 megohm.

Cleaning Sensors, pg. 25).

Resistivity

readings

2. Minimize test sample exposure to air

(ref. Measuring Resistivity, pg. 10).

Unable to calibrate

Film or deposits on electrodes.

Clean cell cup and electrodes (ref.

Conductivity

/

T D S

Cleaning Sensors, pg. 25).

Resistivity

readings much

1. Contamination from previous sample or 1. Rinse cell cup more thoroughly before

lower than expected

from pH sensor well.

measurement. Ensure pH cap is snugly

2. Carbon dioxide in test sample.

in place.

2. See Measuring Resistivity, pg. 10.