Myron L 6P (includes Addendum 10-01) User Manual

Page 8

Advertising
background image

conversion. Generally, using KCl for conductivity, NaCl for resistivity, and
442 (Natural Water characteristic) for TDS will reflect present industry
practice for standardization. This is the setup as shipped from the factory
(ref. Solution Characteristics, pg. 32).
The USER selection allows a custom value to be entered for the
temperature compensation of conductivity and also the conversion ratio if
measuring TDS.

C. Calibration of Each Solution Type

There is a separate calibration for each of the 4 solution types. Note that
calibration of a 442 solution does not affect the calibration of a NaCl
solution. For example: Calibration (ref. Conductivity or TDS Calibration,
pg. 15) is performed separately for each type of solution one wishes to
measure (ref. Conductivity/TDS Standard Solutions, pg. 28).

D. Procedure to Select a Solution

N O T E

: Check display to see if solution displayed (KCl, NaCl, 442 or

USER) is already the type desired. If not:

1.

Press ,

or to select the parameter on which

you wish to change the
solution type.

2.

Press and hold key

about 3 seconds to make

S E L

” appear (see Figure 1).

(For demonstration purposes,
all 4 solution types are shown simultaneously.)

3.

Use

or

key to obtain type of solution desired (ref.

Solution Characteristics, pg. 32). The selected solution type will
be displayed: KCl, NaCl, 442 or User.

4.

Press to accept new solution type.

E. Application of USER Solution Type

1. User Programmable Tempco

This feature allows you to change your Ultrameter’s temperature
compensating factor to another factor between 0-9.99%/°C (ref.

12

M S

MR

CAL

MCLR

COND

T D S

CAL

MCLR

R E S

Figure 1

KCl

NaCl

User

4 4 2

Temperature Compensation, pg. 29). This feature does not apply to pH
or ORP.

a.

As in Procedure to Select a Solution, pg. 12, select “USER”
mode.

b.

With “USER” mode now selected, press . You may now

adjust a temperature compensation from .00%/°C to 9.99%/°C,

by pressing

or .

See example in Figure 2.

c.

Press twice to skip

calibration adjustment and
accept the new tempco (3
times if in TDS mode). You are now ready to measure samples
with your new temperature compensation factor.

2. Disabling Temperature Compensation

a.

As in Procedure to Select a Solution, pg. 12, select “USER”
mode.

b.

With “USER” selected, press

. If the display does not

show .00%/°C, hold

long enough to bring the tempco to

.00%/°C (see Fig. 3).

c.

Press twice (3 times

if in TDS mode). Temperature
compensation is now disabled
(=0) for measurements in
USER mode.

3. User Programmable Conductivity to TDS Ratio

This feature allows you to select a custom conductivity to TDS conversion
ratio for USER mode measurements.

For example: The conversion ratio range is 0.20-7.99 (ie., if conductivity

is 100 µS and TDS is 75 ppm, you would adjust to 0.75) (ref. Conductivity
Conversion to TDS, pg. 32).

13

CAL

MCLR

CAL

MCLR

MR

M S

MR

CAL

MCLR

CAL

MCLR

Figure 3

COND

°C

% /

User

Figure 2

COND

°C

% /

User

Advertising