Myron L 6P (includes Addendum 10-01) User Manual

Page 3

Advertising
background image

FEATURES and SPECIFICATIONS

A. Features:

• Superior resolution 4 digit LCD displays full 9999 µS/ppm.
• Accuracy of ±1% of reading (not merely full scale).
• All electrodes are internal for maximum protection.
• Latest 4 electrode cell technology.
• Waterproof to 3 feet/1 meter.
• Autoranging conductivity/TDS/resistivity.
• Prompts for easy pH calibration.
• Memory saves 20 readings.
• Factory calibrations stored in microprocessor.
• 3 conductivity/TDS solution conversions preprogrammed into

microprocessor.

• USER feature allows:

Programming your own cond/TDS conversion factor.
Programming your own temperature compensation factor.
Disabling temperature compensation.

B. General Specifications

Display

4 Digit LCD

Dimensions (LxWxH)

7.7x2.7x2.5 in.
196x68x64 mm

Weight

13.5oz./383g

Case Material

VALOX*

Cond/Res/TDS Cell Material

VALOX*

Cond/Res/TDS Cell Capacity

0.2 oz./5 ml

pH/ORP Sensor Well Capacity

0.04 oz./1.2 ml

Power

9V Alkaline Battery

Battery Life

>100 Hours/5000 Readings

Operating/Storage Temperature

32-132°F/0-55°C

Protection Ratings

IP67/NEMA 6 (waterproof to
3 feet/1 meter)

* ™ GE Corp.

Additional information available on our web site at:

www.myronl.com

2

C. Specification Chart

If, in the opinion of the factory, failure was due to materials or
workmanship, repair or replacement will be made without charge. A
reasonable service charge will be made for diagnosis or repairs due to
normal wear, abuse or tampering. This warranty is limited to the repair or

replacement of the Ultrameter only. The Myron L Company assumes no

other responsibility or liability.

E.

Ultrameter Models

Resistivity

0-9999 ppm

10-200 ppt

in 5 autoranges

0-9999 µS

10-200 mS

in 5 autoranges

10K

- 30M

ohms

±1% of reading*

±1% of reading*

±1% of reading**

0.01 (<100 ppm)
0.1 (<1000 ppm)
1.0 (>1000 ppm)

0.01 (<100 µS)
0.1 (<1000 µS)
1.0 (>1000 µS)

0.01 (<100K

ohm

)

0.1 (<1000K

ohm

)

1.0 (>1MΩ

ohm

)

0.1 °C/F

±.01 pH

±1 mV

±0.1 °C

0-71°C

32 - 160 °F

0 - 9.99%/°C

Auto

Temperature

Compensation

Adjustable

Temperature

Compensation

Cond/TDS

Ratios

Preprogrammed

Resolution

Adjustable

Cond/TDS Ratio

Factor

0.20 - 7.99

KCl, NaCl, 442™

Ranges

0-14 pH

±999 mV

Accuracy

±.01 pH

±1 mV

0-71°C

32 - 160 °F

0 - 9.99%/°C

0 - 9.99%/°C

0-71°C

32 - 160 °F

0-71°C

32 - 160 °F

0-71°C

32 - 160 °F

Conductivity

pH

ORP

TDS

Temperature

6P

3

ULTRAMETER MODELS

3P

4P

6P

PARAMETERS

pH/ORP/Temp. Conductivity/TDS Conductivity/TDS/pH

Resistivity/Temp. Resistivity/ORP/Temp.

MYRON L COMPANY

2450

Impala

D

rive

Carlsbad, CA 920

1

0

USA
760-438-2021

* up to 100mS/ppt; 100 - 200mS/ppt: ± 1 - 2% typ.

**10K

ohms

- 10M

ohms

D. Warranty/Service

All Myron L Ultrameters have a

2

year warranty except for pH sensors

which have a 6-month limited warranty. If an instrument fails to operate
properly, see the Troublshooting Chart, pg. 27. The battery and pH/ORP
sensor are user replaceable. For other service, return the instrument
prepaid to the Myron L Company.

Advertising