Myron L 6P (includes Addendum 10-01) User Manual

Page 25

Advertising
background image

II.

USER MODE CALIBRATION LINK

A new function has been added to the Ultrameter that makes calibration
of the unit while in "User" mode easier, and more repeatable and accurate
than other calibration methods. It is recommended that this calibration
method be used to provide the highest degree of confidence when the
Ultrameter is used in "User" mode.

A.

Calibration of Ultrameter for use in User Mode

1.

Press the

or

key to power up the Ultrameter.

2.

Calibrate the unit using a Standard Solution. Refer to
CALIBRATION in Operation Manual.

3.

Place the Ultrameter in User mode. Refer to SOLUTION
SELECTION in Operation Manual.

4.

Verify/Set the calibration link. (See below – Setting User Mode
Calibration Link).

B.

Setting User Mode Calibration "Link"

The link function sets or "links" the calibration gain factor of a Standard
Solution to the User solution mode. Once set, the "link" will stay intact
with future calibrations unless the link has been canceled. For more
information on canceling the User Mode Calibration Link refer to the
section below "Canceling User Mode Calibration Link". Follow the steps

below to set either the KCl, NaCl or 442 calibration factor to the User

solution mode.

1.

Press measurement key desired to be “Linked”, i.e.,

,

or

.

2.

Place the Ultrameter in User mode. Refer to SOLUTION
SELECTION in Operation
Manual for selecting the User
Mode.

3.

Press the

arrow key until

the menu "

Linc

" appears.

Fig. 6.

Figure 6

MR

COND

T D S

R E S

COND

T D S

3

4.

Press the

key. The

instrument will display "

S E L

"

and the “

User

” Icon. Fig. 7.

Any additional display of KCl, NaCl or 442 icons indicates a link between
the User solution and the other solution displayed.

5.

Press the

keys to

select a Standard Solution to be
linked to the User mode
calibration constant.
(see Fig. 8,

User

linked to

KCl

).

N o t e

: If none of the Solution Selection icons are displayed (i.e. KCl,

NaCl or 442) nothing has been linked to User mode.

6.

Press the

key to accept the setting. Pressing any of the

measurement keys will exit without changing the setting. User
mode "link" is now complete. The User mode will now use the
calibration gain constant used for the calibration of the Standard
Solution as outlined above.

C.

Canceling User Mode Calibration "Link"

The Ultrameter must be in

User

linked mode in order to cancel the "link".

Refer to SOLUTION SELECTION in Operation Manual.

1.

Press “Linked” measurement key , or

.

Two solution icons will be shown the left side of display - “

User

and another, i.e., “

KCl

”.

2.

Press the

key until the menu "

Linc

" appears. Fig. 5.

3.

Press the

key, the unit will display "

S E L

" and the “

User

Icon.

T D S

COND

R E S

4

MR

CAL

MCLR

CAL

MCLR

M S

MR

CAL

MCLR

Figure 7

User

Figure 8

KCl

User