Myron L 6P and 4P User Manual

Page 14

Advertising
background image

10

4.

Take reading. A display of [- - - -] indicates an overrange

condition.

B. Measuring Resistivity

Resistivity is for low conductivity solutions. In a cell cup the value may

drift from trace contaminants or absorption from atmospheric gasses, so

measuring a flowing sample is recommended.

1.

Ensure pH protective cap is secure to avoid contamination.

2.

Hold instrument at 30° angle (cup sloping downward).

3.

Let sample flow continuously into conductivity cell with no

aeration.

4.

Press

key; use best reading.

NOTE: If reading is lower than 10 kilohms display will be dashes:

[ - - - - ]. Use Conductivity.

C. Measuring pH (6P)

1.

Remove protective cap by squeezing its sides and pulling up.

2.

Rinse sensor well 3 times with sample to be measured. Shake

out each sample to remove any residual liquid.

3.

Refill both sensor wells with sample.

4.

Press

.

5.

Note value displayed.

6.

IMPORTANT: After use, fill pH/ORP sensor well with Myron L

pH Sensor Storage Solution and replace protective cap.

If Myron L pH Sensor Storage Solution is unavailable, use a

strong KCl solution, a pH 4 buffer, or a saturated solution of

table salt and tap water (ref. Cleaning Sensors, 2. pH/ORP, pg.

32).

Do not allow pH/ORP sensor to dry out.

D. Measuring ORP (6P)

1.

Remove protective cap by squeezing its sides and pulling up.

2.

Rinse sensor well 3 times with sample to be measured. Shake

out each sample to remove any residual liquid.

Advertising