Myron L 6P and 4P User Manual

Page 20

Advertising
background image

16

a.

Rinse conductivity cell three times with your standard.

b.

Refill conductivity cell with same standard.

c.

Press

or

, then press

twice in COND/three

times in TDS. The “CAL” icon will appear on the display.

d.

Press

or

to step the displayed value toward the

standard’s value or hold a key down to scroll rapidly through

the reading.

e.

Press

once to confirm new value and end the

calibration sequence for this particular solution type.

3. Resistivity Calibration

Resistivity is the reciprocal of Conductivity. Resistivity is calibrated only

if conductivity is calibrated for the same solution type.

4. Reloading Factory Calibration (Cond or TDS)

If calibration is suspect or known to be incorrect, and no standard

solution is available, the calibration value can be replaced with the

original factory value for that solution. This “FAC” value is the same for

all Ultrameter IIs, and returns you to a known state without solution in

the cell. The “FAC” internal electronics calibration (which bypasses the

electrodes and cell) is not intended to replace calibration with conductivity

standard solutions. If another solution type requires resetting, change

solution type and repeat this procedure.

a.

Press

or

.

b.

Press

. (If in USER solution mode, press CAL key

twice if in Conductivity, and three times if in TDS to skip over

tempco and ratio adjustments.)

c.

Press

key until “FAC” appears and release.

d.

Press

to accept the factory calibration setting.

Advertising