Myron L 6P and 4P User Manual

Page 31

Advertising
background image

27

2.

Press

to display the stored memory records.

3.

Press

repeatedly until you pass the CELL ch location.

The display will show “Auto oFF” (see Figure 24).

4.

Press

to initiate. CAL

will be displayed along with

the “15 SEC” (see Figure

25).

5.

Press

or

to

change the time (see

Figure 26). Maximum

time is shown.

6.

Press

to accept the

change (new time).

XV.

USER MODE CALIBRATION LINC™ FUNCTION

Linc™ function allows easy calibration when in User mode and the user

does not have a user standard solution to calibrate the instrument. This

function will ensure more repeatable and accurate measurements than

many other calibration methods. It is recommended that this function be

used to provide the highest degree of confidence when the Ultrameter II

is used in “User” mode. When Linc is used, the User mode is linked to

another standard, i.e. if User and KCl are linked, a KCI standard solution

is used to calibrate the instrument. It is that simple.

A. Calibration of Ultrameter II for use in User Mode

1.

Press

or

key.

2.

Calibrate the unit using a Standard Solution, (ref.

CALIBRATION, pg. 14).

3.

Place the Ultrameter II in User mode, (ref. SOLUTION

SELECTION, pg. 11).

Figure 25

CAL

Figure 26

CAL