Myron L 6P and 4P User Manual

Page 33

Advertising
background image

29

6.

Press

key to accept the setting. Pressing any of the

measurement keys will exit without changing the setting. User

mode “Linc” is now complete. The User mode will now use the

calibration gain constant used for the calibration of the

Standard Solution as outlined above.

C. Canceling User Mode Calibration “Linc”

The Ultrameter II must be in User linked mode in order to cancel the

“Linc”, (ref. SOLUTION SELECTION, pg. 11).

1.

Press “Linked” measurement key

,

or .

Two solution icons will be shown in the left side of display -

User” and another, i.e., “KCl”.

2.

Press

key until the menu “Linc” appears, (see

Figure 27).

3.

Press

key, the instrument will display both “SEL” and

the “User” Icon.

4.

Press

key until “User” is the only solution icon being

displayed.

5.

Press

key.

6.

The User mode calibration “Linc” has now been canceled.

NOTES:

1. To maintain repeatability, use the same standard solutions for future

calibrations.

2. Calibration of the Ultrameter II Gain Factor for User mode is not

available when the calibration linc has been established. The other

calibration functions (i.e. Temperature Compensation %/C settings and

TDS Ratio settings) are still intact. To perform a calibration of the User

mode as described in User Calibration Conductivity/TDS, pg. 15, the

User Mode Linc should be canceled. See above Canceling User Mode

calibration “Linc”.

3. Once a “Linc” has been established for User mode, the “Linc” will apply

Advertising