33

Wear protective clothing, eye protection and gloves! Follow all safety
instructions and observe limitations of use and operating limitations
(pages 28-30).

Check the discharge tube for a tight fit.
After two days tighten up the locking
nut(s).

Note:

Warning:

2.

Mounting the

discharge tube

Before mounting the discharge tube
make sure that the seal is inserted.

seripettor

®

:

sealing ring, discharge valve, discharge
tube (fig. 1).
seripettor

®

pro:

sealing ring, adapter, gasket, discharge
tube (fig. 1').

1.

Check seals

Is everything in the package? Confirm that your package includes:

seripettor

®

:

Bottle-top dispenser seripettor

®

,

discharge tube, filling tube, spare dispensing cartridge,

bottle adapters (PP) and this operating manual.

seripettor

®

pro:

Bottle-top dispenser seripettor

®

pro,

discharge tube, filling tube, spare dispensing cartridge, adapter for

discharge tube and filling valve, mounting tool, bottle adapters (PP) and this operating manual.

Assembly

Adapters for bottle thread

Filling tube length

seripettor

®

GL 32-33, S 40

250 mm

seripettor

®

pro

GL 32-33, S 40

125 - 240 mm

seripettor

®

:

Slide the discharge tube onto the valve
block and firmly finger-tighten the locking
nut (fig. 2).
seripettor

®

pro:

Slide the discharge tube onto the adapter
and firmly finger-tighten the locking nut.
Then, slide the adapter onto the valve
block and firmly finger-tighten the locking
nut (fig. 2').