Error Limits

Error limits related to the nominal capacity (= maximum volume)
indicated on the instrument, obtained when instrument and distilled
water are equilibrated at ambient temperature (20 °C/68 °F). Testing
takes place with a completely filled instrument and with uniform and
smooth dispensing.

Nominal volume
ml

A*

±

%

µl

CV*

%

µl

2

1.2

24

0,2

4

10

1.2

120

0,2

20

25

1.2

300

0,2

50

* A = Accuracy, CV = Coefficient of Variation

Error limits seripettor

®

!

20 °C

Ex

38

e.g.

Volume

A*

±

%

µl

CV*

%

µl

V

N

10

1.2

120

0,2

20

V

T

= 50% N

5

2.4

120

0,4

20

V

T

= 10% N

1

12

120

2,0

20

Note:

The maximum error limit for a single measurement can be calculated
EL = A + 2 CV (e.g. for volume 10 ml: 120 µl + 2 x 20 µl = 160 µl).

Partial volume
The percentage values for A and CV are relative to the nominal volu-
me (V

N

) and must be converted for partial volumes (V

T

).

V

N

V

T

A

T

= · A

N

* A = Accuracy, CV = Coefficient of Variation