Dispensing – BrandTech Scientific seripettor User Manual

Page 36

Advertising
background image

c

a

!

a) Hold the discharge tube and loosen the

stopper cap/screw cap.

b) Hold the discharge tube orifice on the

inner wall of a suitable receiving vessel.

c) Manually press the pump assembly

all the way down and hold it there
(see fig.).

d) Put the receiving vessel beneath the

discharge tube orifice.

e) Loosen the volume selector thumb

screw one-half turn, set the pointer
to the desired volume and then re-
tighten it (see fig.).

2.

Dispensing

Dispensing

a) Hold the discharge tube and loosen

the stopper cap/screw cap.

b) Hold the discharge tube orifice on the

inner wall of a suitable receiving ves-
sel.

c) Press the piston down slowly and

steadily without using excessive force
to the lower stop. Then let it rise up
slowly (fig. c).

d) Wipe off the discharge tube against

the inner wall of the receiving vessel.

e) Reattach the stopper cap to the

discharge tube.

Wear protective clothing, eye protection and gloves! Liquid may accu-
mulate in the stopper cap. To avoid splashes dispense slowly. Follow
all safety instructions and observe limitations of use and operating
limitations (pages 28-30).

Warning:

1.

Setting the volume

36

Note:

When setting the volume, liquid will be dispensed!

Advertising