Troubleshooting – BrandTech Scientific seripettor User Manual

Page 48

Advertising
background image

Problem

Possible cause

Corrective action

Troubleshooting

Piston difficult to move

Formation of crystals

Replace dispensing cartridge (page 43).

Dispensing of incompatible
liquid

Check for operating exclusions
('Dispenser selection chart', page 31).

Liquid above piston seal

Dispensing cartridge worn

Replace dispensing cartridge (page 43).

Filling not possible

Valve not correctly mounted

Mount valve correctly (pages 41, 42).

Volume adjusted to minimum
setting

Set to required volume (page 36).

Sticking filling valve

Clean filling valve (pages 41, 42). If nec-
essary, replace the valve and sealing ring
(observe operating exclusions, page 30).

Dispensing cartridge worn

Replace dispensing cartridge (page 43).

Dispensing not possible

Sticking discharge valve

Clean discharge valve. If necessary replace
discharge valve.

Air bubbles in the
instrument

Reagent with high vapor pres-
sure has been drawn in too
quickly

Slowly draw in reagent.

The instrument has not
been primed

Prime the instrument (page 35).

Dispensing cartridge worn

Replace dispensing cartridge (page 43).

Filling tube not firmly con-
nected or damaged

Cleaning procedure (page 40). Push the filling
tube on firmly. If necessary, cut off approx.
1 cm of tube at the upper end and re-connect
it or replace filling tube.

Valves not firmly connected,
contaminated or damaged

Cleaning procedure (page 40). Tighten fill-
ing valve and discharge tube. If necessary,
replace valves and seals.

Dispensed volume
is too low

Discharge tube not firmly
connected or damaged

Push the discharge tube on firmly. Replace a
deformed or damaged discharge tube.

Filling tube not firmly
connected or damaged

Cleaning procedure (page 40). Push the filling
tube on firmly. If necessary, cut off approx.
1 cm of the tube at the upper end and re-
connect it or replace filling tube.

Filling valve not firmly
connected or damaged

Cleaning procedure (page 40). Tighten filling
valve. If necessary, replace valve and sealing
ring.

Frequent dispensing > 40 ºC Replace dispensing cartridge (page 43).

48