Min. and max. temperature

2)

Pointer

In this menu item, the device status is displayed. If the sensor detects
a fault, "OK" will be displayed. If a fault is detected, a flashing failure
message is outputted sensor-specifically, e.g. "E013". The fault is also
displayed in clear text, e.g. "No measurement value".

Information:
The fault message as well as the clear text indication are also carried
out in the measured value display.

Meas. reliability

Sensor status

Available with the trend curve:

l

"X

-Zoom": Resolution in minutes, hours or days

l

"S

top/Start": Interrupt a recording or start a new recording

l

"U

nzoom": Reset the resolution to minutes

Trend curve

Menu section, service

In this menu item you simulate a user-defined level or pressure value
via the current output. This allows you to test the signal path, e.g.
through connected indicating instruments or the input card of the
control system.

The following simulation variables are available:

l

Percent

l

Current

l

Pressure

With Profibus PA sensors, the selection of the simulated value is made
via the "Channel" in the menu "Basic adjustments".

How to start the simulation:

2)

Temperature: -50 … +150 °C.

Sensor status

Curve presentation

Simulation of
measured val-
ues

14

Indicating and adjustment module for IPT-1* sensors