Level, Offset, Ext gating input – BNC 630 User Manual

Page 44: Dualtone examples

Advertising
background image

39

6. Level

In this field enter the output level, from 4 mVp-p to 10 Vp-p in 1 mV steps or from -44.0 dBm to +21.0 dBm in .1 dBm steps.

NOTE: The level specified is a 50 ohm LOADED level. This is the level of the signal which will appear across a 50 ohm load
connected to the SIG Out connector. Into an open circuit, the output swing will be twice the value entered.

Offset

You can enter an offset voltage for the output waveform. For more information on output offsets refer to section 4.4.

Ext Gating Input

(rear panel connector)

This TTL compatible input can be used to gate the output signal on or off. A logic high voltage turns off the output.
For further information on the Ext Gating Input, refer to section 2.0.

Dualtone Examples

The following table lists the parameters for some common dualtones used in the American Telephone Network.

Name

Frequency 1

Frequency 2

On Time (Ton)

Off Time (Toff)

Dial Tone

350 Hz

440 Hz

1 mS

0 mS

Ringback

440 Hz

480 Hz

2,000 mS

4,000 mS

Busy

480 Hz

620 Hz

500 mS

500 mS

Reorder

480 Hz

620 Hz

250 mS

250 mS

DSN Preempt

440 Hz

620 Hz

1 mS

0 mS

European

420 Hz

440 Hz

various

various

Note: The Phase Offset field value may be set to any value too properly generate these tones.