7.6 Pulse Generator Mode

1. Positive Only

This field, when set to “Y”, does not let the output signal go below 0V. This feature is handy when driving circuitry that cannot
accept negative voltages. This feature eliminates having to readjust the offset voltage to obtain a positive only signal every time the
output level is changed. When this field is set to “Y”, the output level is halved, so that the output signal swings from 0V to ½ the
specified level into a 50 ohm load (or 0V to the specified level into an open circuit).

NOTE: If you are using the Pulse Mode to drive digital logic, you should use the SYNC Out connector on the front panel. This
signal has the same shape as the SIG Out signal, but comes from a TTL/CMOS driver.

2. Duty Cycle

In this field enter the duty cycle for the pulse waveform, from 0% (always low) to 100% (always high) in 1% steps. The duty cycle
is the percentage of the waveform period that the output is high. A waveform with a 10% duty cycle for example means that the
output is high for 10% of the time and low for the remaining 90% of the waveform period.

3. Continuous / Triggered mode

In this field select whether the waveform is generated continuously or on a single event basis. If you select Continuous mode, then
the waveform is immediately restarted once the end of the waveform is reached. If you select a Triggered mode, then the waveform
is halted after the last point is reached, and the waveform is not restarted again until another trigger occurs. The trigger may come
simultaneously from three sources:

1. Pressing the

Trigger

key

2. Applying a low-to-high transition on the Ext. Trig In connector
3. Sending an ASCII "T" to the RS232 port

To set Continuous mode, press

1

. To set Triggered mode, press

0

. Pressing arrow keys or rotating the wheel will toggle the mode.

4. Repetition Frequency

In this field enter the frequency at which the function will be repeated. You may enter from 0 Hz (DC) to 2.000000 MHz in 1 Hz
steps. If a value of 0 Hz is entered, the function waveform clock is halted. A high going synchronization pulse is given on the SYNC
Out connector (front panel) at the start of the waveform.

In Triggered mode, the waveform is generated once per trigger event. The duration of each waveform will be 1/Repetition Freq.

Tip

: Very low rep rate frequencies (<< 1 Hz) can be generated by setting the Repetition Frequency to 1 Hz. (This fills the waveform

memory with 16,000 points of the desired function.) Then go to Arbitrary Waveform Mode and enter the Clock Frequency to obtain
the desired function rep rate. Repetition Rate = Clock Frequency / 16,000

5. Level

In this field enter the loaded output level, from 4 mVp-p to 10 Vp-p in 1 mV steps. Into an open circuit, the voltage swing will be

twice the specified peak-to-peak value.

Waveforms with very narrow pulses may have signal energy above the cutoff of the output filter (10 MHz). The resultant signal
may have a distorted shape and reduced peak-to-peak output swing.

If the Pos Only? field is set to “Y”, then the output level is halved, so that the output signal swings from 0V to ½ the specified level
into a 50 ohm load (or 0V to the specified level into an open circuit).

Offset

You can enter a DC offset voltage by pressing the Offset key. For more information see section 4.4.

Pulse Gen

Pos Only? N Duty Cycle: 50 %