Biological Shakers Innova® 44/44R — Operating manual

Abb. 18:

Rear Panel

Fig. 18:

Rear Panel

8.5

Optional gassing manifold kit

This option is factory-installed. The manifold delivers gas into the chamber and ultimately into the
flasks themselves via twelve ports. You may elect to use splitters after the manifold to increase
the number of ports for your application.

8.6

Optional UV germicidal lamp

This option, a germicidal ultraviolet lamp, is placed inside the Service Drawer, outside the
chamber to help reduce the risk of contamination. The lamp is identified on the display screen as
UV.
The UV germicidal lamp is factory-installed and available on refrigerated units only.

8.7

Optional photosynthetic lamps

This factory-installed option (for refrigerated units only) provides nine photosynthetic growth
lamps inside the chamber. They can be turned on and off manually, by the easily set
programmable timer, or by computer. These lamps are identified on the display screen as GRO.
The replacement bulb part number is P0300-0221.
The recommended operating temperature when using this option is 15 °C – 37 °C.
The maximum operating temperature when using photosynthetic lamps is 70 °C.

8.8

Optional humidity monitor

This optional factory-installed accessory allows you to monitor actual humidity levels in the
chamber throughout your run. With humidity monitor, the maximum temperature is 60 °C.

8.9

Fill the water reservoir

If you choose to use the spill pan/reservoir as a water reservoir to reduce evaporation:

1. Temporarily remove the four thumbscrews (with their washers) that secure the front grille in

place (Fig. 19 on p. 36).

1 Adjustable Foot

2 Air Intake Vents (Do not block)

3 Power Connector

1

2

3

WARNING!

Personnel injury and equipment damage!

Never try to operate the UV Germicidal Lamp while the shaker door is open.

8