3 setup screen, Setup screen, En (see setup screen on p. 40) – Eppendorf Biological Shakers Innova 44R User Manual

Page 40: Summ th 16:18

Advertising
background image

Biological Shakers Innova® 44/44R — Operating manual

40

Abb. 23:

Summary Screen

Fig. 23:

Summary Screen

The only elements you can modify in this screen are setpoints. To change setpoints in this
screen
:

1. Turn the Knob until the desired setpoint is highlighted, then click the Knob in.

The setpoint will begin to flash.

2. Turn the Knob to the right to increase the number, or to the left to decrease it. One click left or

right will increase the setpoint by an increment of one (one whole unit or one tenth unit,
depending on the parameter). Move the Knob more rapidly (you can spin it) to change the
value by larger increments.

3. Click the Knob in to set and save the new value.

4. Repeat the above steps to change any or all of the other setpoints.
To move out of this screen and into the next:

1. Use the Control Knob to highlight SUMM, then click the Knob in.

SUMM begins to flash.

2. Turn the Knob to the right until the next screen, Setup (SET), appears. If you turn too far and

enter another screen, just turn the Knob back to the left to recapture the SET screen.

3. Click the Knob in to select the screen and to work in it.

9.4.3

Setup screen

Here you can set the day of the week and the time (on a 24-hour clock). This screen also allows
you to lock all of your settings from further changes, and to mute or enable the audible alarm.

PARAM

ACTUAL

SET

RPM

100

100

ºC

45.1

45.0

HRS

0.0

0.0

%RH

50.0

N/A

SUMM

Th 16:18

Hint!

The current day (Su, Mo, Tu, We, Th, Fr or Sa) and time always remains visible in the lower
righthand corner.

Hint!

If you highlight an item, change it, but do not save your selection, after a few seconds the screen
will revert to its previous setting.

9