5 profile control commands, Profile control commands, Appendix a: remote programming – Eppendorf Biological Shakers Innova 44R User Manual

Page 67

Advertising
background image

67

Biological Shakers Innova® 44/44R — Operating manual

15.5

Profile control commands

For example, the following PW command:
PW<Space>25.0<Space>150<Space>0.0<Space>2<Space>30<Space>0<Space>1<CR>
will set Program 1, Step 1 to the following conditions:

Temperature setpoint 25.0 °C

Agitation setpoint 150 rpm

CO2 setpoint 0.0 %

Step Time 2 hours 30 minutes

UV Lamp OFF

GRO Lamp ON

Code

Meaning

Format

PC

Clear Profile (clears profile
step)

PC<Space><Arg1><Space><Arg2><CR>
where Arg1= Profile number (range 1 to 4) and Arg2=Step
number (range 1 to 15)
NOTE: If Arg2 is not included, all 15 steps of the selected
profile will be cleared.

PM

Profile Mode (returns status of
current profile run/hold mode)

PM<CR> There are no parameters to this command.
Return data format:
<Arg1><Space><Arg2><Space><Arg3><CR>
where Arg1=Run/hold status (0 = end, 1 = run), Arg2=Profile
number and Arg3=Step number
NOTE: If a profile has not been selected to run, a 0 will be
returned for Profile number and Step number.

PR

Read Profile step in memory

PR<Space><Arg1><Space><Arg2><CR> where Arg1=
Profile number (range 1 to 4) and Arg2=Step number (range
1 to 15)
Return data format:
<Arg1><Space><Arg2><Space><Arg3><Space>...<CR><L
F>
where Arg1=Profile number (range 1 to 4), Arg2=Step
number, Arg3=Temperature setpoint, Arg4=Agitation
setpoint, Arg5=CO2 setpoint (future option, reads 0.0),
Arg6=Step hours, Arg7=Step minutes, Arg8=UV lamp, and
Arg9=Grow lamp

PS

Profile Stop/Start (stops or
starts a program profile)

PS<CR> With no other parameters, any profile currently
running will be stopped.
PS<Space><Arg1><Space><Arg2><CR>
where Arg1=Profile number (range 1 to 4), and Arg2=Step
number (range 1 to 15).
NOTE: If Arg2 is not included, the profile will start in Step 1.

PW

Write Profile step to memory

PW<Space><Arg1><Space><Arg2><Space><Arg3><Spac
e>...<CR>
where Arg1=Profile number (range 1 to 4), Arg2=Step
number (range 1 to 15), Arg3=Temperature setpoint,
Arg4=Agitation setpoint, Arg5=CO2 setpoint (future option);
Enter 0.0, Arg6=Step hours (range 0 to 99), Arg7=Step
minutes (range 0 to 59), Arg8=UV lamp (0 = off, 1 = on), and
Arg9=Grow lamp (0 = off, 1 = on)

15

Appendix A: Remote programming