Biological Shakers Innova® 44/44R — Operating manual

set, use the Control Knob to select the field, click the Knob to select the field (it will flash), turn
the Knob to select On, then click the Knob in to save the setting. In our sample Step 1
(see Fig. 31 on p. 47), however, GRO remains Off.

11. To program Step 2 (see Fig. 31 on p. 47): use the Control Knob to highlight the arrows at the

bottom left of the screen. Click the Knob to make the arrows flash, then turn the Knob to the
right until Step 2 appears. Click the Knob in to work in this screen, and repeat Steps 3 – 10.
Whatever step is shown on the right-hand side of the display is settable ((see Fig. 31 on
p. 47), it
is Step 2). To scroll between steps, select the arrows (bottom left), then turn the
Control Knob clockwise or counterclockwise, then select the desired step.

Abb. 31:

Program 1, Step 2

Fig. 31:

Program 1, Step 2

12. Continue to program as many as 15 steps in the same manner. For our sample program, we

have only three steps (see Fig. 32 on p. 47).

Abb. 32:

Program 1, Step 3

Fig. 32:

Program 1, Step 3

Hint!

DO NOT select “Save” yet!

1 Step Number

PRG1 - STEP 01

02

Time

08:00 03:00

ºC

37.0 40.0

RPM

150 225

UV

Off

Off

GRO

Off Off

Save Cancel

1

Hint!

The time entered for each step is for that step only; it is not cumulative (i.e. the elapsed time from
the start of the program).

1 Step Number

PRG1 - STEP 02

03

Time

03:00 12:00

ºC

40.0 4.0

RPM

225 0

UV

Off

Off

GRO

Off Off

Save Cancel