Det-Tronics PW9200x Pathwatch IR Gas Detection System User Manual

Page 23

Advertising
background image

SYMPTOM: Voltage at NUMERATOR (blue) test

jack relative to GROUND (green) is not within
0.3 to 0.5 vdc after proper alignment and out-
put zero adjustment.

POSSIBLE CAUSE: Improper preamplifier gain

setting for path length.
REMEDIAL ACTION: See Step 14 of “Optical

Alignment/Signal Strength” procedure.

SYMPTOM: RATIO is noisy or erratic.

POSSIBLE CAUSE: Instrument not optically

aligned.
CHECK: Voltage at NUMERATOR test jack.

Voltage should be 0.3 to 0.5 vdc.

REMEDIAL ACTION: If NUMERATOR voltage

is low, realign as described in “Optical
Alignment” procedure.

SYMPTOM: RATIO voltage is negative or 4 to 20

ma output is below 4 ma.

POSSIBLE CAUSE: The ZERO adjustment poten-

tiometer located on the front panel is not set
properly.
CHECK: Setting of ZERO potentiometer.
REMEDIAL ACTION: Adjust ZERO poten-

tiometer until RATIO output is 0.1 ± 0.1
vdc.

POSSIBLE CAUSE: Instrument malfunction.

CHECK: NUMERATOR and DENOMINATOR

voltages when a bare flashlight bulb or
bare 60 watt light bulb is held in front of
receiver window.

REMEDIAL ACTION: Replace each printed

circuit board (one at a time) and recheck
voltages. Contact factory if fault cannot
be located.

SYMPTOM: RATIO voltage is steady and does not

appear to be responding to heavy hydrocar-
bons within the monitored path.

POSSIBLE CAUSE: Malfunction of signal pro-

cessing circuitry.
CHECK: Response of RATIO when a piece of

plastic film (roughly 8 mils thick) is placed
in front of the receiver window. The
RATIO voltage should increase to at least
+2 vdc.

REMEDIAL ACTION: Substitute spare boards

and assemblies until defective part is
located and RATIO responds properly to
plastic film. Defective part can be
returned to factory for analysis and repair.

POSSIBLE CAUSE: Object in beam.

REMEDIAL ACTION: Remove obstruction.

SYMPTOM: HC Alarm does not activate when

hydrocarbon vapors are present in monitored
path, or HC Alarm activates too often.

POSSIBLE CAUSE: Improper HC Alarm threshold

setting.
CHECK: Voltage at HC ALARM SETPOINT

jack relative to GROUND. Refer to
Calibration graph in APPENDIX for con-
centration indicated by alarm threshold
setting.

REMEDIAL ACTION: Adjust setpoint if

required. See “Hydrocarbon Vapor Alarm
Threshold Adjust-ment” procedure.

POSSIBLE CAUSE: Malfunction of HC Alarm cir-

cuitry.
CHECK: RATIO voltage relative to voltage on

HC ALARM SETPOINT jack. If RATIO
voltage exceeds alarm setpoint and alarm
does not activate, a malfunction of alarm
circuitry exists.

REMEDIAL ACTION: Replace rear panel p.c.

board and HC alarm relay on
connector/relay board.

SYMPTOM: RATIO output obtained during calibra-

tion check is not within the range indicated by
the calibration graph.

POSSIBLE CAUSE: Excessive hydrocarbon

background at installation site.
CHECK: Hydrocarbon background along

instrument path length with an indepen-
dent hydrocarbon vapor analyzer.

REMEDIAL ACTION: Wait until hydrocarbon

background along path length of instru-
ment is low, then recheck calibration.

POSSIBLE CAUSE: Calibration shift after factory

calibration due to some parameter within the
instrument, or to some change in the installa-
tion parameters.
CHECK: Calibration to be sure that calibra-

tion check was performed properly and
calibration point obtained is repeatable.

REMEDIAL ACTION: If calibration check is

repeatable and not within calibration
range, adjust CALIBRATION ADJUST-
MENT potentiometer on front panel with
calibration tube filled and installed.

21

95-8400

Advertising