Barco MD-4221 User Manual

Page 12

Advertising
background image

12

RF Emissions
CISPR 11

Class B

Harmonic emissions
IEC 61000-3-2

D

Voltage fluctuations
IEC 61000-3-3

Complies

including domestic
establishments and those
directly connected to the
public low-voltage power
supply network that
supplies buildings used for
domestic purposes

2. Guidance and manufacturer's declaration - electromagnetic immunity

The MD-4221 is intended for use in the electromagnetic environment specified below.
The customer or the user of the MD-4221 should assure that it is used in such an
environment.

Immunity test

IEC 60601 Test level

Compliance Electromagnetic

environment
-guidance

Electrostatic
discharge(ESD)

IEC61000-4-2

6kV Contact

8kV air

6kV Contact

8kV air

Floors should be
wood, concrete or
ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%

Electrical fast
transient/burst

IEC 61000-4-4

2kV for power
supply lines

1kV for
input/output lines

2kV for power
supply lines

1kV for
input/output lines

Mains power
quality should be
that of a typical
commercial or
hospital
environment

Harmonic emissions
IEC 61000-3-2

D

Voltage fluctuations
IEC 61000-3-3

Complies