Menu Structure and Navigation When Using One Sensor

Nutshell description

Mode

Menu item

Function description

1x

SEnA unIT

Select the unit of measurement of channel A

2x

SEnA TAR

Set or clear the zero point of channel A

3x

SP

Setting aid for mechanical pressure switches

4x

VErS

Display of the unit’s software version

5x

Back to measured value display

Detailed description

Mode Press

1x

Selecting the unit of

measurement

(channel A)

Use the arrow keys (◄, ►) to
choose between the units of
measurement offered:
Pressure:

bar, PSI, MPA

Temperature: °C, K, °F
Flow rate:

l/m (in), g/m (in)

Mode Press

2x

Setting or clearing the zero point

(channel A)

Use the arrow keys (◄, ►) to
choose between SET and CLR
(see Setting the Zero Point,
starting on page 13).