JULABO SW23 Shaking Water Baths User Manual

Page 28

Advertising
background image

Menu functions

28

8.4. ATC - Absolute Temperature Calibration

Internal sensor (T

F

)


Measuring point(T

M

)

ATC serves to compensate a temperature difference that might
occur between circulator and a defined measuring point in the bath
tank because of physical properties.

The difference temperature is determined (

T= T

M

- T

T

)

and stored as correcting factor (example

T = -0.2 °C).

1.

Press the cursor key

and enter

at the same time.

2.

Use the cursor keys

to select the menu option

„At0“.

3.

With the edit keys

select "At1" and then press

enter

.





Measuring point

Enter the corrective value.

4.

Using the cursor keys

and the edit keys

set the correcting factor (example -0.20 °C) and then

press enter

.

5.

Press

and

at the same time.

The temperature on the measuring point rises to a temperature of
37.0 °C and is indicated on the
MULTI-DISPLAY (LED).

The ATC function stays activated until resetting to 00.0 °C.

Recommendation:
Use a calibrated temperature measuring instrument.