JULABO SW23 Shaking Water Baths User Manual

Page 33

Advertising
background image

Shaking Water Baths

33

Disturbances that are not indicated.

Warning:

D a n g e r o f e l e c t r i c s h o c k !
The exchanging of the fuses are to be carried out only by qualified

service personnel.

 Before exchanging the fuses, turn off the mains power switch and disconnect

the power plug from the mains socket!

 Only use fine fuses with a nominal value as specified.

Mains fuses

 Pull the mains plug from the power outlet before opening the

equipment!

 The mains fuses are located behind side plate (13).

The side plate is fastend to the casing with 6 screws .
(Fine fuse T 10.0 A, dia. 5 x 20 mm)

Sicherungen
Fuses
Fusibles secteur

Pump motor overload protection

The pump motor is protected against overloading. After a short
cooling interval, the motor will automatically start running.