34

10.1. Acoustical signals and their differentiation

Three different signals are generated by the sound
generator as follows:
- an alarm signal
- a warning signal

- a time signal

The signals can easily be recognized and differentiated, even from
a good distance. Required actions can be initiated immediately.



 The Alarm signal is a continuous sound signal.

Heater and circulator pump (SW23 only) are completely and

permanently switched off

.

(see page 31)

......

 The warning signal is issued at regular intervals

(signal - pause - signal - pause).

The actual bath temperature is higher than the set overtemperature

value or lower than the set undertemperature value

(see page 22)

..

 The time signal is issued in the intervals

(double signal - pause - double signal - pause).

Countdown of the adjusted operating time commences to zero,

after which a time signal is issued at intervals

(see page 25)