31

8.7. Evaluation of the temperature limits





The high- and low-temperature limit can be adjusted in two ways
(see page 22)

3. As pure warning function with an acoustic warning signal in

regular intervals.
Adjustment „Li 0“ – factory adjustment

4. As temperature limit by switching-off the heating. Adustment „Li

1“
The alarm is indicated by optical and audible signals
(continuous tone) and on the MULTI-DISPLAY (LED) appears
the error message "Error 01".

1.

Press

and

at the same time.

2.

Use the cursor keys

to select the menu option

Limit. (example: „Li 0“).

3.

Select the alternative state with the edit keys

and

confirm the selection with the ENTER key

(display „Li 1“).

4.

Press

and

at the same time.


9. Safety installation (with shutdown function)



(excess temperature protection)

These safety installations is independent of the control circuit.
When the temperature of the bath liquid has reached the safety
temperature, a complete shutdown of the heater and pump (only
CW-models) is effected.

The alarm is indicated by optical and audible signals (continuous
tone) and on the MULTI-DISPLAY (LED) appears the error
message "Error 01".