KD Scientific Legato 200 Series User Manual

Page 25

Advertising
background image

25

Configuring the KDS Legato 200 Series

K

D

S

L

e

g

a

to

2

0

0

S

e

ri

e

s

U

s

e

r’s

M
a

n

u

a

l

KD Scientific Legato 200 Series Syringe Pump User’s Manual

Publication 5617-002-Rev F

Setting the Force Limits

The maximum linear force of the syringe pump is 75 lbs

or 35 kgs. This force is sufficient to damage delicate, low

volume syringes, as well as many standard syringes if

pumping viscous fluids. The % of maximum force applied

by the pusher block can be changed.

In the set up screen, press

Force

.

Set the force level.

Press

Other Value

to enter a custom force %

Accept

the changes made by pressing

and go back to the Set Force screen.

Cancel

the changes made by pressing

and go back to the Set Force screen.

For a Custom Force %

Enter the precentage for custom force %

Accept

the changes made by pressing

and go back to the Set Force screen.

Cancel

the changes made by pressing

and go back to the Set Force screen.

NOTE: A minimum value of approximately

30% is required to overcome friction forces

in the mechanism. You may have to set the

force level to a higher value depending on

the backpressure/friction in your system.

Press Here

Press here to enter

custom force %

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

Suggested Force Level Settings for Common Syringes

Syringe Type/Material

Capacity

Force Setting

Common Manufacturer(s)

Plastic Syringes

≤20ml

50%

BD plastic, Sherwood,

Airtite, Terumo

>20ml

100%

Stainless Steel

All

100%

KDS, Cole-Parmer

Glass/Glass

≤20ml

20%

KDS Glass, Cadence (Popper)

>20ml

30%

Glass/Plastic

≤1000µl

30%

Hamilton, SGE

≤5ml

50%

>5ml

100%

Advertising